Re: TOCSY

charlie thibault (cth@bruker.com)
Tue, 7 Sep 93 9:36:40 EDT

According to Graham Barlow:
>
> Does anyone know where I can get a TOCSY pulse program for an MSL type
> pulse programmer ie. a .PC file?
>
> Graham
> ._____________________________________________________________________.
> | Dr. Graham Barlow | On the VAX mail: gkb1 |
> | NMR Service | On Tower mail: gkb1 |
> | Dept. of Chemistry | JANET mail: gkb1@UK.AC.YORK |
> | Univ. of York, UK | Internet mail: gkb1@tower.york.ac.uk |
> +----------------------+----------------------------------------------+
>
The following pulse program was submitted by Dr. Stefan Steuernagel
of Bruker Analytische Messtechnik.

-- 
==================================================================
	
       ...     ...	Charles G. Thibault
      .    * .    .	Manager of Software Development
       .   . .   .	Bruker Instruments Inc.
       B R U K E R	Billerica, MA. 01821 USA
       .   . .   .	(508)667-9580 x218
      .    * .    .	cth@bii.bruker.com		Internet
       ...     ...     ...!uunet!bii!cth		UUCP

================================================================= ==============================8<------------------------------ ; MLEV17PH.PC ; TOCSY/HOHAHA EXPERIMENT WITH MLEV-17 CYCLE FOR MIXING ; SEE A. BAX & D.G. DAVIS, J. MAGN. RESON. 65, 355 (1985) ; PHASE SENSITIVE USING DIGITAL PHASE SHIFTER ; BUFFERED ACQUISITION

DE=DW*3/4 ; CALCULATE DEADTIME DELAY PROT F1 XT ; PROTECT F2 AND XT OUTPUTS VC1=0 ; PRESET COUNTER FOR EVOLUTION DELAY VC2=DS+NS ; PRESET COUNTER FOR AQ WITH DS

START, 10U 30U [DECPL=S1] LOOP VC2 TIMES 10U [PH2 @TPPI RGATE] 2U D11 [F2 @PLS1 RGATE] ; PROTON 90 DEG. PULSE 2U [PH2 0 RGATE] 2U LOOP VC1 TIMES DW [RGATE] ; EVOLUTION END LOOP 30U [DECPL=S2 RGATE] 2U INCLUDE MLEV17F2.INCL ; SET MLEV17 PARMS. D3 [STA RGATE] D0 ; ACQUISITION 10U END LOOP 10U 2U [++VC1] ; INCREMENT COUNTER FOR ; EVOLUTION DELAY 2U [++TPPI] ; INCREMENT LIST FOR ; DIGITAL PHASE SETTING 2U [VC2=NS] ; RESET COUNTER FOR AQ ; WITHOUT DUMMY SCANS 10U GOTO START

BEGIN LISTS TPPI, 0 90 180 270 PLS1, +X -X -X +X +Y -Y -Y +Y PLS2, +X +X -X -X +Y +Y -Y -Y PLS3, +Y +Y -Y -Y -X -X +X +X PLS4, -X -X +X +X -Y -Y +Y +Y PLS5, -Y -Y +Y +Y +X +X -X -X RLS, +X -X -X +X +Y -Y -Y +Y END LISTS

; FOR MLEV-17 PARAMETERS SEE ALSO MLEV17F1.INCL ; SET DP TO GET 90 DEG. PULSE OF CA. 25 US TO YIELD SPIN LOCK FIELD ; OF ABOUT 10 KHZ ; USE O2 FOR CARRIER FREQUENCY, REC=F2 ; DECOUPLER MODE: DO ; D11 = 90 DEGREE PULSE WIDTH ; D12 = 180 DEGREE PULSE WIDTH ; D21 = LOW LEVEL 90 DEGREE PULSE ; D22 = LOW LEVEL 180 DEGREE PULSE ; D13 = 2.5 MS, TRIMPULSE ; ; MIXING-TIME ; C1 = ------------- , ; 66 * D21 ; ; FOR DIRECT CONNECTIVITIES USE CA. 40 MS MIXING TIME ; FOR REMOTE CONNECTIVITIES USE CA. 80 MS MIXING TIME ; NS=8*N ; DS=8 ; ; FOR PROCESSING USE: ; MC2=W, ND4=2, REV=N, WDW1=WDW2=Q, SSB1=SSB2=2 OR 3

hear starts the INCLUDE-file: ; MLEV17 FOR TOCSY/HOHAHA EXPERIMENTS INCLUDING TRIM PULSES ; FOR INSTRUMENT SETUP IN INVERSE MODE ; INCLUDE SEQUENCE FOR: ; MLEV17F2.PC ; INVMLEVF.PC

D13 [F2 @PLS3 RGATE] LOOP C1 TIMES ;SET C1 FOR MIXING TIME ;CYCLE IS 66*D21 D21 [F2 @PLS2 RGATE] D22 [F2 @PLS3 RGATE] D21 [F2 @PLS2 RGATE] D21 [F2 @PLS4 RGATE] D22 [F2 @PLS5 RGATE] D21 [F2 @PLS4 RGATE] D21 [F2 @PLS4 RGATE] D22 [F2 @PLS5 RGATE] D21 [F2 @PLS4 RGATE] D21 [F2 @PLS2 RGATE] D22 [F2 @PLS3 RGATE] D21 [F2 @PLS2 RGATE] D21 [F2 @PLS4 RGATE] D22 [F2 @PLS5 RGATE] D21 [F2 @PLS4 RGATE] D21 [F2 @PLS4 RGATE] D22 [F2 @PLS5 RGATE] D21 [F2 @PLS4 RGATE] D21 [F2 @PLS2 RGATE] D22 [F2 @PLS3 RGATE] D21 [F2 @PLS2 RGATE] D21 [F2 @PLS2 RGATE] D22 [F2 @PLS3 RGATE] D21 [F2 @PLS2 RGATE] D21 [F2 @PLS4 RGATE] D22 [F2 @PLS5 RGATE] D21 [F2 @PLS4 RGATE] D21 [F2 @PLS2 RGATE] D22 [F2 @PLS3 RGATE] D21 [F2 @PLS2 RGATE] D21 [F2 @PLS2 RGATE] D22 [F2 @PLS3 RGATE] D21 [F2 @PLS2 RGATE] D21 [F2 @PLS4 RGATE] D22 [F2 @PLS5 RGATE] D21 [F2 @PLS4 RGATE] D21 [F2 @PLS2 RGATE] D22 [F2 @PLS3 RGATE] D21 [F2 @PLS2 RGATE] D21 [F2 @PLS2 RGATE] D22 [F2 @PLS3 RGATE] D21 [F2 @PLS2 RGATE] D21 [F2 @PLS4 RGATE] D22 [F2 @PLS5 RGATE] D21 [F2 @PLS4 RGATE] D21 [F2 @PLS4 RGATE] D22 [F2 @PLS5 RGATE] D21 [F2 @PLS4 RGATE] D22 [F2 @PLS3 RGATE] ; PURGE 180 DEGREE PULSE END LOOP D13 [F2 @PLS3 RGATE] ; TRIM PULSE

==============================8<------------------------------

-- 
!!! please notice the new ZIP-code !!!
=====================================================================
       ...     ...      Dr. Stefan Steuernagel
      .    * .    .     Bruker Analytische Messtechnik GmbH
       .   . .   .      Silberstreifen
       B R U K E R      D-76287 Rheinstetten
       .   . .   .      Voice: +49-721-5161-124
      .    * .    .     Fax:   +49-721-517101
       ...     ...      EMail: ste%bamdemo%bamsoft@Germany.EU.net
=====================================================================