Re: FREQ. LISTS on MSL

charlie thibault (cth@bruker.com)
Tue, 22 Feb 94 9:20:41 EST

According to Lazaros Kakalis:
> I am trying to set up an X nucleus saturation transfer pulse program
> on an MSL. I am usingan automation program to vary the irradiation period
> and I would like to include an F1 frequency list with my irradiation points.
> My program compiles OK and the first offset is read correctly but
> the second one is either erratically misset or the CANNOT SYNTHESIZE FREQUENCY
> message is given. Can anybody help? Thanks.

The frequency lists in AU are actually lists of offsets to F1 not
absolute frequencies. If an absolute frequency is given the
CANNOT SYNTHESIZE FREQUENCY message will result. If this is not
the case e-mail me the AU program, the version of DISMSL and
the frequency list used and I will help you find the problem.

-- 
==================================================================
	
       ...     ...	Charles G. Thibault
      .    * .    .	Manager of Software Development
       .   . .   .	Bruker Instruments Inc.
       B R U K E R	Billerica, MA. 01821 USA
       .   . .   .	(508)667-9580 x218
      .    * .    .	cth@bii.bruker.com		Internet
       ...     ...     ...!uunet!bii!cth		UUCP

=================================================================