Re: TCFLUSH and Kaeppeli's procedure

John Grundy (ram!jeg)
Fri, 6 Dec 1996 14:33:25 -0500

Hi David,

The messages you describe are normal on systems equipped with the
SCM and BSN18 type shim controls. The TCFLUSH and Kaeppeli's
procedures are not needed for this hardware. They may be applied
to the newer BSMS type shim controls. If you need further assistance,
please contact your local service office.

Best Regards,
John Grundy

>
> Dear Bumlist Readers,
>
> We are using new software from Bruker on our AMX600. Lately, I have been
> writing simple SIMPLEX based tune programs to speed up shimming. When I
> give the tune command, the system responds with the warning:
>
> No TCFLUSH; No Kaeppeli's procedure.
>
> Am I missing something? What are TCFLUSH and Kaeppeli's procedure? Are
> these things that we ought to have?
>
> Sincerely,
>
> David Naugler
> IMBB
> SFU
>
>

-- 