Dear all,

first of all I'd like to thank to all who responded to my question about
sofrtware attenuation of X channels on our AMX spectrometer. Almost all
of you suggested to use dhi/dlo command. Of course, that's the first and
obvious solution. However, there was a minor problem with our
spectrometer. When I turned the decoupler on by dlo command I lost any
signal (without decoupling everything was OK). Then I realized that a
high pass filter in the proton channel between the probe and preamp
might help. It really did and now everything is OK.

Richard

--
* * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * *
*                                                                 *