> My question (for XWINNMR users) is:
> Is it possible to change PARMODE (from, say, 1D to 2D) from within
> automation program, which runs a series of experiments?
> The problem is that when you do it in "eda", XWINNMR takes care
> about fetching "required additional parameter files" from
> the directories "/u/exp/stan/nmr/par/standard2D" or "standard3D".
> Is there any way to fetch these files (which files?) explicitly in AU?
Use RPAR instead:
GETCURDATA
RPAR("standard2D","all");
QUIT
The result will be the same
-- :::: Old E-Mail: gsc@bruker.de is now: Guenter.Schmidt@Bruker.DE ::::: Dr. Guenter Schmidt | Bruker Analytik GmbH | Phone: +49 7243-504-443 -Software Department- | Rudolf Plank-Str. 23 | Fax: +49 7243-504-480 | D-76275 Ettlingen |