Re: AQ solved

From: Dr. Gerhard Eber (Gerhard.Eber@bruker.de)
Date: Fri Aug 04 2000 - 00:35:57 PDT


The exact formula is:

AQ = (TD + extraDwell(digtyp))/(2*SW*SFO1)

Of course, SW*SFO1=SW_p, but this is correct only after the acquisition
is finished and processing has been done once.

extraDwell is a number between 0 and 3 which depends on the digitizer
(HADC+ = 1 or 3, FADC = 2, IADC = 4, others = 0) and describes the number of
dwells which must be fed to the digitizer, before the first real point
comes out from the pipeline.

Gerhard Eber
_____________________________________________________________
Dr. Gerhard Eber Phone +49-721-5161450
Software Department FAX +49-721-5161480
Bruker Analytik GmbH Email: Gerhard.Eber@bruker.de
D-76287 Rheinstetten Am Silberstreifen
_____________________________________________________________