Hi all,
Finally I realized that my problem is caused by the wrong setting of little
delta, which should be double the gradient pulse! Under xwinnmr 2.6, there
is no au program called setdiffparm to transfer the acquisition parameters
to dosy processing parameters, so one has to check these parameters with
great caution.
All the inputs regarding my problem will be appreciated!