- Originally, CDCl3 appears at 7.32 ppm in peak pick output (pps, pp).
- Got to calibrate, set it to 7.25ppm.
- Pps (or pp) STILL shows CDCl3 at 7.32 ppm
- Now, use 'efp'. Peak pick output is now correct.
We are using uxnmr version 930301.
Rudi Nunlist