Re: peak picking in UXNMR

joe edward meier (jem@bruker.com)
Wed, 27 Oct 93 16:52:29 EDT

The problem you are experiencing with peak values not matching what
is displayed is most-likely due to CY not being set to a value. Make
sure to do a "defplt" and then to set "CY" to a value, it is sometimes
set to zero. Then make sure "MAXI" and "MI" are set acorrding to
the new CY value. You can display the y-axis to make this easier to
see.

Joe

-- 
==================================================================
    Joseph Meier, Ph.D.
    Senior NMR Software Scientist

Bruker Instruments Inc. USA Billerica, MA. 01821

Phone: 508-667-9580 Fax: 508-663-9177

jem@bii.bruker.com Internet !{uunet,ulowell}!bii!jem UUCP =================================================================