AMX Receiver Gating

Udo Guenther (ug%bamsoft.uucp@Germany.EU.net)
Fri, 21 Jan 94 9:09:01 MEZ

RE: AMX receiver gating

The reiver handling in the current UXNMR versions for AMX
(930601) is as follows:

The DE Delay is calculated acoording to DE=1.427*DW in quad mode.
Upon execution, DE is splitted in 2 parts:
4usec + (DE - 4usec).

After 4usec, the receiver is opened and the phase is reset.
After DE, the digitizer is started. The receiver is opened
earlier in order to get the first point right.

Due to the fact that the pulse is delayed by 2 usec, the 4 usec
are actually squeezed to 2usec.

Please contact our applications group (Dr. W. Bermel
Email: ber%bamdemo%bamsoft@Germany.EU.net or
Dr. R. Kerssebaum Email:rke%bamdemo%bamsoft@Germany.EU.net)
who developed this procedure, if you need more
information, e.g. on dependencies of particular hardware
configurations.

Bruno Guigas
Software Dept.
Bruker Karlsruhe