Problem with digital to ana

Sergio Scrofani (Sergio_Scrofani@muwayf.unimelb.edu.au)
Wed, 25 Oct 1995 17:43:01 +1000

Subject: Time: 3:49 PM
OFFICE MEMO Problem with digital to analog? Date: 25/10/95

Dear BUMmers,

If you have to port data Avance data to non-XWIN software packages, the
following may be useful.

I've found that converting digitally oversampled Avance data to the analogue
form using "convdta" leads to an incorrectly calculated TD in acqus - 122
points (the magic number for oversampled data) less than it should be.
Although slices in F2 appear fine in 2D experiments, it wreaks havoc when the
F1 transform is applied. Setting TD to the correct value overcomes this
problem.

Regards,
Serge

**********************************************************
Dr. Sergio D. B. Scrofani Email: sergio_scrofani@muwayf.unimelb.edu.au
School of Chemistry Tel: +61-3-9344-6498
University of Melbourne Fax: +61-3-9347-5180
Parkville, 3058
Victoria, Australia
**********************************************************