DPX artifacts

Reiner DIEDEN (DIEDEN@MVA1.noh.be.solvay.com)
Fri, 21 Mar 1997 11:18:00 +0000 (GMT)

Hello everybody,

Thank's to everybody who helped us with our problem concerning the artifacts on
the DPX300.

Most people suspected theseartifacts to be decoupling artifacts - and were
completely right !
Thanks to the very kind help of Mr. Kaufer from Bruker - Switzerland (special
thanks to him !), we could eliminate other possibilities (talk-through of
decoupler- or lock-channel, ..) and finally found that without decoupling or
CW, there were no artifacts.
So, I tried different cpd-decoupling schemes (waltz16, bb, garp, mlev16 and
dipsi2), and found that dipsi-2 works best in our conditions.

Thanks again,

Reiner
Reiner
-------------------------------------------------------------------------------
Reiner Dieden
Laboratoire de RMN
AN003/R37
Solvay
310, rue de Ransbeek
B-1120 Brussels
Belgium
tel : +32 (0)2/264.26.37
fax : +32 (0)2/264.20.55
e-mail : reiner.dieden@solvay.com
-------------------------------------------------------------------------------