Thank you very much Abil, Andy, Bill, Jan, Jose, Kurt, Mike, Nils,
Peter, Robert, Ronnie, Shan, William, and Yuri!
Some of you suggested using macro "humpcal" for chloroform lineshape
measurement. Unfortunately, it will not work for XWINNMR 1.3 and we
cannot afford hardware modification required for significant software
upgrade.
I was able to determine these linewidths manually using the procedure
described by several people:
"Expand the peak such that it almost fills the screen. Put the cursor
onto the spectrum by clicking on the spectrum. Move the mouse cursor to
a position from where you want to measure the distance and click the
middle mouse button. A marker is going to appear at the position where
you clicked. Now move the mouse to another location. You will see that
the info window updates the mouse position and also shows the distance
in Hz."
Thanks again,
Piotr
Piotr Dobrowolski
NMR/EPR Manager
Department of Chemistry and Biochemistry
Utah State University
0300 Old Main Hill, Logan UT 84322-0300
telephone: (435) 797-1673
fax: (435) 797-3390
This archive was generated by hypermail 2b29 : Tue Dec 30 2003 - 01:04:22 PST