RE: vt shim cooling

From: Andrew Gibbs (andrew.gibbs@bruker.co.uk)
Date: Thu May 22 2008 - 03:45:46 PDT


Hi,

I assume this question has already been answered, but I just thought I'd throw in my comments for good measure!

1) There is a temperature sensor in the shimset so if you have a concern you can check this value and take appropriate action! The example au program in the standard release is named 'coiltemp', which runs in the background and logs the temperature to a file. For extra safety you could also think to modify this program to take action if a temperature range is exceeded (for example of this try ftp://ftp.bruker.co.uk/pub/uknmr/auprogs/coiltempmon).

2) If you need one an adaptor is available for Bruker spectrometers that allows the connection of gas to flow over the shimset. You should contact your local Bruker office to get more details and a price!

Best wishes,

Andrew

Dr Andrew Gibbs Phone (+44) 024 7685 5200
Team Leader, NMR Applications Fax (+44) 024 7646 5317
Bruker BioSpin Limited email: andrew.gibbs@bruker.co.uk
Banner Lane
Coventry
CV4 9GH
U.K.

Registered in England and Wales with registration no 923986
Registered office is Banner Lane Coventry CV4 9GH

> -----Original Message-----
> From: ligy [mailto:ligy@mail.sioc.ac.cn]
> Sent: 16 May 2008 18:03
> To: BUM
> Subject: vt shim cooling
>
>
> Dear all,
>
> Varian machines have a cooling system to protect the room
> temperature shim
> unit, which is very useful for a long vt experiment. I dont
> notice such a
> device on our Bruker av500. Is there a similiar cooling
> system on Brukers
> for vt experiments? And what is your solution in practice?
>
> Thank you in advance!
>
> Guangyu
>
> ~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~
> Guangyu Li
> Manager, Associate Professor
> NMR Facility
> Shanghai Institute of Organic Chemistry
> Chinese Academy of Sciences
> 354 Fenglin Lu
> Shanghai 200032 China
> Telephone: 86-21-54925012
> Fax: 86-21-54925039
>
> ~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~
> ______________________________________________________________
> ____________
>
> To subscribe/unsubscribe: Please send a note to
>
> bruker-users-request (at) purcell(dot)cchem(dot)berkeley(dot)edu
>
> For subscriptions, please include a list of Bruker instruments you are
> using, see http://www.cchem.berkeley.edu/nmr/BUM.txt
> E-mail archives are at
http://calmarc3.cchem.berkeley.edu/archives/bum.html
___________________________________________________________________________

________________________________________________________________________
This email has been scanned by MessageLabs Email Security System on behalf
of Bruker BioSpin Limited.
________________________________________________________________________



This archive was generated by hypermail 2b29 : Tue Dec 23 2008 - 07:12:18 PST